Table 6-9 Measurement specifications (continued)
Resistance
• Measurement method
• Open-circuit voltage
• Zeroing error
• Input protection
• Response time
Diode/Continuity
• Measurement method
• Test current
• Open-circuit voltage
• Continuity threshold
• Continuity level
• Audible tone
• Input protection
U3401A User's and Service Guide
2-wire Ohms
Limited to +6.0 VDC
0.05 Ω or less (excluding test lead resistances) in each range when Rel operation is used
500 V DC or AC RMS
Approximately 1.5 seconds for 5 MΩ and ranges below 5 MΩ.
0.83 mA ± 0.2% constant current source, open-circuit voltage limited to < 5 V
Approximately 0.5 mADC
Limited to < + .0 VDC
10 Ω fixed
Approximately below +50 m VDC
Continuous beep for continuity and single tone for normal forward-biased diode or
semiconductor junction.
500 V DC or AC RMS
Specifications and Characteristics
6
93