Table 4-4 2-Wire Ω Verification Test; Resistance Verification Test - Agilent Technologies U3401A User's Manual And Service Manual

41/2 digit dual display multimeter
Hide thumbs Also See for U3401A:
Table of Contents

Advertisement

4
Performance Test

Resistance Verification Test

Configuration: 2- Wire Ω
1 Select the resistance function.
2 Select each range in the order shown in
Table 4-4 2-wire Ω verification test
Notes:
1 Specifications are for 2-wire ohms function using the Rel operation enabled to eliminate lead
64
resistance value indicated. Compare measurement results to the
appropriate test limits shown in
appropriate source settling).
Function
Input
2-wire Ω
0.00 Ω
0.0000 kΩ
0.000 kΩ
0.00 kΩ
0.0000 MΩ
0.000 MΩ
450.00 Ω
4.5000 kΩ
45.000 kΩ
450.00 kΩ
4.5000 MΩ
30.000 MΩ
resistance.
Table 4-
4. Provide the
Table 4-
4. (Be certain to allow for
Range
Error from nominal one year
500 Ω
±50 mΩ
±0.3 Ω
5 kΩ
±3 Ω
50 kΩ
±30 Ω
500 kΩ
±300 Ω
5 MΩ
50 MΩ
±3 kΩ
500 Ω
±500 mΩ
±4.8 Ω
5 kΩ
±48 Ω
50 kΩ
±480 Ω
500 kΩ
5 MΩ
±4.8 kΩ
50 MΩ
±93 kΩ
U3401A User's and Service Guide
[1]
[1]
[1]
[1]