4
Performance Test
Calibration Overview
Make sure you have read
N O T E
instrument.
Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your local Agilent
Service Center. The U3401A is supported on automated calibration
systems at Agilent service centres only.
Calibration Interval
A one- year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at regular
calibration intervals. Accuracy specifications are not warranted beyond the
one- year calibration interval. Agilent does not recommend extending
calibration intervals beyond two years for any application. When an
adjustment is required, contact your local Agilent Service Center.
58
"Test Considerations"
on page 60 before calibrating the
U3401A User's and Service Guide