Table 4-2 DC voltage verification test

Set the calibrator output to 0 V before disconnecting it from the multimeter input
C A U T I O N
terminals.

DC Current Verification Test

1 Connect the calibrator to the front panel Hi and Lo input connectors.
2 Select each function and range in the order shown in
3 Make a measurement and observe the result. Compare measurement
62
Function
Input
DC voltage
0.00 V
0.0000 V
0.000 V
0.00 V
0.0 V
450.00 mV
4.5000 V
45.000 V
450.00 V
900.0 V
Provide the input shown in
results to the appropriate test limits shown in
allow for appropriate source settling when using the Fluke 5520A).
Range
500 mV
5 V
50 V
500 V
1000 V
500 mV
5 V
50 V
500 V
1000 V
Table 4-
3.
Table 4-
U3401A User's and Service Guide
Error from nominal one year
±40 μV
±0.4 mV
±4 mV
±40 mV
±400 mV
±130 μV
±1.3 mV
±13 mV
±130 mV
±580 mV