Diode Verification Test

Configuration: Diode
1 Connect the calibrator to the front panel Hi and Lo input terminals.
2 Select each function and range in the order shown in
3 Make a measurement and observe the result. Compare measurement

Table 4-5 Diode verification test

Frequency Verification Test

Configuration: Frequency
1 Select the frequency function.
2 Select each range in the order shown in

Table 4-6 Frequency verification test

U3401A User's and Service Guide
Provide the input shown in
results to the appropriate test limits shown in
allow for appropriate source settling when using the Fluke 5520A).
Function
Voltage
Diode
1.0000 V
2.0000 V
voltage and frequency indicated. Compare measurement results to the
appropriate test limits shown in
appropriate source settling).
Function
Voltage
Frequency
2 V
Table 4-
5.
Range
2.3 V
2.3 V
Table 4-
Table 4-
6. (Be certain to allow for
Input frequency
Range
4.5 kHz
5 kHz
Performance Test
Table 4-
5.
Table 4-
5. (Be certain to
Error from nominal one year
±1 mV
±1.5 mV
6. Provide the input
Error from nominal
one year
±0.7 Hz
4
65