Table 4-3 Dc Current Verification Test - Agilent Technologies U3401A User's Manual And Service Manual

41/2 digit dual display multimeter
Hide thumbs Also See for U3401A:
Table of Contents

Advertisement

Table 4-3 DC current verification test

Connect calibrator to multimeter's A and Lo terminals before applying 10 A.
C A U T I O N
U3401A User's and Service Guide
Function
Input
0.00 μA
DC current
0.0000 mA
0.000 mA
0.00 mA
450.00 μA
4.5000 mA
45.000 mA
450.00 mA
0.0000 A
0.000 A
4.5000 A
9.000 A
Range
Error from nominal one year
500 μA
±0.05 μA
±0.4 μA
5 mA
±4 μA
50 mA
±40 μA
500 mA
500 μA
±0.27 μA
±2.6 μA
5 mA
±26 μA
50 mA
500 mA
±0.26 mA
5 A
±0.5 mA
10 A
±5 mA
5 A
±11.7 mA
10 A
±27 mA