True Rms Ac Current; Table 6-3 Aci Resolution, Full Scale Reading, And Burden Voltage - Agilent Technologies U3401A User's Manual And Service Manual

41/2 digit dual display multimeter
Hide thumbs Also See for U3401A:
Table of Contents

Advertisement

6
Specifications and Characteristics

True RMS AC Current

Table 6-3 ACI resolution, full scale reading, and burden voltage [±(% of reading + count)]
Mode
Range
Resolution
0.01 μA
500 μA
0.0001 mA
5 mA
0.001 mA
50 mA
AC
0.01 A
500 mA
coupling
5 A
0.0001 A
10 A
0.001 A
0.01 μA
500 μA
0.0001 mA
5 mA
0.001 mA
50 mA
AC + DC
0.01 mA
500 mA
coupling
5 A
0.0001 A
10 A
0.001A
Notes:
1 Typical at full range reading and voltage across the input terminals.
2 In 10 A range, > 10-20 Aac is readable for 20 seconds maximum with audio warning.
3 Input Current < 3 A RMS.
4 Input Current > 35 μA RMS.
5 Accuracy specified at input >5% of range and >1 A for 10 A range except other specified.
86
Accuracy (One year; 23°C ± 5°C)
Maximum
30 Hz to 50
reading
Hz
510.00
1.5% + 50
5.1000
1.5% + 40
51.000
1.5% + 40
510.00
1.5% + 40
[3]
5.1000
2% + 40
[2]
[3]
20.000
2% + 40
510.00
5.1000
51.000
510.00
5.1000
[2]
20.000
50 Hz to 2
2 kHz to 5
kHz
kHz
0.5% + 20
1.5% + 50
0.5% + 20
1.5% + 40
0.5% + 20
1.5% + 40
0.5% + 20
1.5% + 40
0.5% + 30
(< 1kHz)
0.5% + 30
(< 1kHz)
0.5% + 30
1.5% + 60
0.5% + 30
1.5% + 50
0.5% + 30
1.5% + 50
0.5% + 30
1.5% + 50
0.5% + 40
(< 1kHz)
0.5% + 40
(< 1kHz)
U3401A User's and Service Guide
[5]
Burden
5 kHz to 20
[1]
voltage
kHz
[4]
< 0.06 V
3% + 75
3% + 60
< 0.6 V
3% + 60
< 0.08 V
3% + 60
< 0.8 V
< 0.3 V
< 0.6 V
[4]
< 0.06 V
3% + 85
3% + 70
< 0.6 V
3% + 70
< 0.08 V
3% + 70
< 0.8 V
< 0.3 V
< 0.6 V

Advertisement

Table of Contents
loading

Table of Contents