C
ALIBRATION
6 Retrieve the printed measurement page from the copier/printer.
7 Click Measure in the Get Measurements pane.
The Measurement Options dialog box appears.
Page Type is automatically set to the type that you selected in
8 Check the settings and click Measure.
The Information dialog box appears.
9 Place the ES-1000 in the calibration cradle.
When you place the ES-1000 in the calibration cradle, white point calibration is used to
calibrate the spectrophotometer and compensate for gradual drifts in the instrument. Make
sure the sample aperture is in full contact with the white tile on the calibration cradle. If
you do not place it correctly in the calibration cradle, the spectrophotometer does not
return accurate measurements.
10 Click OK.
The Measurements dialog box appears. Instructions for measuring the strips appear in the
Status field.