Hioki ST4200 Instruction Manual page 265

Partial discharge detector
Hide thumbs Also See for ST4200:

Advertisement

It is also possible to conduct a system reset with the
List of initial settings
Types of Measurement Modes
Setting item
Measurement mode
Type of PD to be Measured
AC PD Measurement Operation Settings
Symbol
U
Ulimit
Voltage Upper Limit
Ilimit
Voltage Current Limit
f
Tref
Sampling Time Width
Sc
Qsen
Measurement Sensitivity
Qth
AC PD Threshold Value
Er
Qmax Evaluation Rate
fH
High Cutoff Frequency
Default setting
Normal
Setting item
Applied voltage
Voltage Frequency
PD Axis Scale
*RST
communications command.
Initialize Settings
AC
Default
setting
200 V
5000 V
0.2 mA
50 Hz
100 ms
300 pC
x40
10.00 pC
50 pps
1000 kHz
HIOKI ST4200B971-00
2 6 5
Initialize all
Initialize
Initialize
Settings
all
*RST
*RST