4
Perform measurement.
Press the
START
Voltage will automatically be output to the object under measurement from AC power insulation
withstand voltage tester and AC PD measurement will begin.
IMPORTANT
When the trigger mode is
measurement is completed.
When the trigger mode is
key is pressed.
When finished measurement make sure to press the
The voltage value can be compensated for using the vernier function.
See:
Executing Vernier
key.
[Single]
voltage output also automatically ends once a single
[Repeat]
measurement and voltage output continues until the
HIOKI ST4200B971-00
STOP
key.
1 2 6
STOP