Significant ditferences between ohmmeter types are:
(1)
ohmmeters did not differ so much in these respects, the re-
the amount of internal voltage they use (2) the currents they
sistance measurements as given in the table would be more
require to make the meter needle deflect full scale for each
typical.
range, and (3) the scale readings on the meter itself.
If
TABLE
4-3
Diode
A ONLY
B ONLY
ALTER.
CHOP.
D134
Forward
Forward
Forward
Forward
D138
Reverse
Reverse
Reverse
Reverse
D139
Reverse
Reverse
Reverse
Reverse
D234
Forward
Forward
Forward
Forward
D238
Reverse
Reverse
Reverse
Reverse
D239
Reverse
Reverse
Reverse
Reverse
D501
Reverse
Forward
D502
Reverse
Forward
D503
Forward
Reverse
D504
Forward
Reverse
See footnote 1.
See footnote 1.
D562
Forward
Reverse
D572
Reverse
Forward
D566
Reverse
Reverse
foot
.
tnote
2,
D576
Reverse
Reverse
See footnote 2
See footnote
D579
Unbiased
Unbiased
See footnote 3.
See footnote 3.
D580
Unbiased
Unbiased
Unbiased
Unbiased
D581
Forward
Forward
Forward
Reverse
D585
Reverse
Reverse
Slightly reverse biased.
Reverse
See footnote 4.
D586
Reverse
Reverse
Reverse
100 ke-Reverse
D590
Unbiased
Unbiased
See footnote 5.
See footnote 5.
D506
Forward
Reverse
D507
Forward
Reverse
;
D508
Reverse
Forward
See footnote 1.
See footnote 1.
D509
Reverse
Forward
1. Same as A ONLY bias conditions when channel A is electronically
switched
''on";
same as
B ONLY
2, Reverse biased.
Conducts only during duration of positive-going trigger pulses.
3. Unbiased.
Conducts when triggers from Blocking Oscillator stage are applied.
4. Conducts only when the positive-going portion of the alternate trace sync pulse is applied to the diode.
5. Unbiased.
Conducts only during the negative-going backswing as 7590 field collapses.
conditions
when
channel
B
is
4-9