SECTION 2
OPERATING INSTRUCTIONS
FUNCTIONS OF CONTROLS
AND CONNECTORS
Signal input connector for channel A or
or dc
coupling
into
the amplifiers,
or
grounding of Input Attenuators.
Nine-position
siwtch
used
to
select
the
calibrated vertical-deflection factors,
permit the gain of the common output
amplifier to be accurately set. Adjusted to
make the basic deflection factor of channel
A accurate at .1 V/CM
Potentiometer to provide variable uncali-
brated attenuation up to about a 2-to-]
ratio
between
the
calibrated
deflection
factors and up to approximately 100 v/cm
1O0v/em using X10 gain.
Potentiometer to provide for shifting the
Screwdriver-adjustable potentiometer to set
the trace to coincide with the horizontal
[X-axis] center-line when the GAIN switch
is set to X1IO and
VERTICAL POSITION
adjustment after adjusting the Xl
POSI-
TION RANGE control first.)
Screwdriver-adjustable potentiometer to set
the trace to coincide with the Type 82
Two-position switch to provide operational
in-phase (NORM, or out-of-phase (INV.)
output for channel A,
Two-position switch to provide operational
in-phase [NORM]
or out-of-phase (INV.}
Two-position switch to select either XI or
X10 gain for both channels simultaneously.
When switch is in the black X1
position,
When the switch is in the blue X10 position,
read the blue scale of the V/CM switches
Four-position switch to allow channel A
to be used independently, to synchronize
the switching with the oscilloscope's sweep,
to provide for switching the two channels
vp
if
positive
volfages
ore
applied
or
if corresponding
VERTICAL
POSITION control
is
turned
trace
moves
down
if
negative
voltages
are
applied
or
if
VERTICAL
POSITION
control
is
turned
counterclockwise.
In
INWERT.
position,
trace
movement
is
epposite
to
that
described
for
the
NORM.
position.
at one of two preselected chopping rates,
or to allow channel B to be used indepen-
dently.
Holds the Type 82 securely in the oscillo-
scope plug-in compartment.
(Located at
bottom center on the front panel.)
Securing Rod
CHOP RATE
Two-position slide switch provides the fol-
(see Fig. 2-1]
lowing
chopped-mode — switching-rate
First Time Operation
To get acquainted with the Type 82 for the first time,
the following operating procedure is suggested:
1, Set the internal CHOP RATE slide switch |see Fig. 2-1)
to the 100 KC position,
2. Insert the Type 82 into a Type 580-Series Oscilloscope,
tighten the Securing Rod, and turn the power on.
3. Allow about 2 to 3 minutes warm-up time and free run
the ascillascope sweep at |
millisec/cm.
4, Set the Type 82 front-panel controls [both channels) as
follows:
AC-.DC-GND
DC
VICM
]
VARIABLE VOLTS/CM
CAL.
A POLARITY
NORM.
B POLARITY
NORM,
VERTICAL POSITION
Centered
GAIN
XI
MODE
A ONLY
Fig.
2-1.
Right
side
front
portion
view
of
the
Type
8&2
showing
the location of the CHOP RATE switch.
2-1