Using the same oscilloscope sweep rates as those used
to obtain the oscillograms shown in Fig. 5-14 a, b and c,
examine the waveform displayed on your oscilloscope crt.
For a waveform 2 centimeters in amplitude, distortion should
not exceed
—-1 millimeter, If the first 100-nsec-duration por-
tion of the front corner of the waveform is not within
-1
millimeter of being square, adjust C417, C418, C456, C477",
C487 and R492** (see Fig. 5-13} for optimum squere corner
and amplitude level, The portion of the waveform that each
of these adjustments affects is given in Table 5-5,
Approx. Time Duration (as measured
Adjustment
from the front leading corner)
C417
4 nsec
C418
60 nsec
;
C456
8 nsec
Cé77*
100 nsec
C487
Ty
6 nsec
R492**
Affects al! 100 nsec
29. Channel A X10 High-Frequency
Disconnect the output end of the 50-ohm termination from
INPUT B ond connect the output to INPUT A connector.
Set
the MODE switch to
A ONLY
If the waveform display does not have a square corner
which
is within
--1
millimeter as shown
in Fig.
5-14 a, b
and ¢
(based on a waveform amplitude of 2 centimeters),
adjust C316, C318, C328, C356, C377*, C387 and R392**
(see Fig. 5-13) for optimum square corner and amplitude
level.
Use various sweep rates to examine the
waveform
from time to time to obtain best overall results,
The Chan-
nel A high-frequency adjustments have the same effect on
the
waveform
as
the
corresponding
adjustments did
for
After completing
this
step turn off
the calibrator.
Dis-
connect the various items connected to the Type 82 and the
oscilloscope.
*Added in Type §2
Plug-in
Units
$/N 390 and up.
**Added
in
Type
§2
Plug-In
Units
SN
361D-up.