Chapter 11: Qualification Reports
The Time of Sterilization calculates the total time of sterilization by combining the
Steam quality criterion, along with the measured reference temperature. Time is
counted if the measured reference temperature is at or above the process temperature
and the steam quality as defined by Temperature minus calculated Saturation
Temperature (T-Sat) is within the specified range.
Figure 112: Time of Sterilization – Definition of Limits
Select Back to return to the Edit Parameters screen.
If Edits are completed or no edits were required, select the Next arrow to proceed.
Kaye Validator AVS User Manual
161