Chapter 11: Qualification Reports
With all edits to the Qual / Exposure times complete, select Define Additional Cycles.
Figure 107: Define Additional Cycles screen
The Define Add'L Cycles screen is utilized if additional cycles, other than Qual and
Exposure, need to be defined. If no additional cycles need to be defined, select Report
Parameters at the top of the screen to proceed to the next step. The system allows up
to 15 additional cycles to be defined. Select the
button at the top right of the screen
to navigate to the Cycles window. The method and functionality for defining cycles is
the same as defined in Exposure Start / End.
In the Cycles window all created cycles are displayed with the defined Start and End
times. The Cycles window also allows modifications of cycles names. UnMark Cycle is
used to delete a cycle.
Figure 108: Define Additional Cycles Screen
Kaye Validator AVS User Manual
157
Need help?
Do you have a question about the Validator AVS and is the answer not in the manual?