Chapter 6. Defining Study Setups
It is also possible during report generation to define events to monitor during the study.
These events are listed in the reports and can be used to define reporting tool cycles.
When finished, press Calibration Parameters to enter the qualification parameters.

6.8. Specifying Calibration Parameters

Figure 64: Calibration Parameters Screen
The Validator AVS provides both sensor calibration and verification.
Before running a qualification study, regulations require performing sensor calibration
to maximize accuracy and ensure that the system meets required uncertainty criteria.
During sensor calibration, the Validator AVS automatically compares each sensor to
the IRTD (NIST-traceable temperature standard). The variation for each sensor at the
low and high calibration point is stored in the SIM as an offset. Offsets are then applied
to each sensor reading during the Qualification study. You can specify a two-point
calibration, or a two-point calibration with one check point.
Sensors that do not meet the User defined uncalibrated sensor criteria are marked as
failed and are not calibrated. Calibration and verification temperatures, chosen set
points, stability criteria, and deviation criteria all constitute a calibration signature.