7.2. Loading a Setup into the Validator AVS
IMPORTANT: When you load a setup with a different calibration signature (calibration
parameters and/or temperature units) than the setup currently in
memory, any sensors already calibrated are marked as uncalibrated. If
you reload the original setup, and you have not calibrated sensors
using the new setup, the offsets are restored, and the sensors marked
as calibrated.
To load a setup:
From the Main Menu go to Asset Details and then select a setup to be used for
your Qualification study, from the list of available setups. Buttons for initiating a
study are active and pressing Initiate Calibration or Initiate Verification starts
the corresponding process.
After pressing Discover, available devices are displayed. Select the Validator AVS to
be used for your Qualification study. Docking mode is preselected (highlighted in
orange), but it is possible to select any Kaye Validators AVS that is currently available
through a network connection.
Select the desired Validator AVS and press Connect. The Validator AVS then loads
the setup and reads sensor information. This process may take a few seconds to
complete. When completed, the Start Calibration screen displays.
Kaye Validator AVS User Manual
Figure 66: Initiate a Calibration from Asset Setups
Chapter 7. Calibrating and Verifying Sensors
95
Need help?
Do you have a question about the Validator AVS and is the answer not in the manual?