Interval Calculations View
The Interval calculations are available as a live screen if group specific interval
calculations are specified in the setup.
The live screen data for the interval calculations is calculated on the display rate. The
display rate is a fixed update rate for the refresh of all live screens. It is five seconds for
SIM slots 1-3 and six seconds if SIM slot 4 is used in the setup.
Interval reports are generated on the storage data rate, which can be specified in the
setup separately.
Note: Live screen interval calculations and the interval calculations report may show
discrepancies if the storage rate is different from the display rate. To avoid such
discrepancies, it is required to set the storage rate to the display rate.
Kaye Validator AVS User Manual
Figure 83: Interval Calculations Live View Screen
Chapter 8. Qualification Study
114