Measurement
–20°C (lower limit of the setting range) to 20°C (upper limit of the setting range)
range
–40°F (lower limit of the setting range) to 40°F (upper limit of the setting range)
Setting Setting unit
range
Ther-
K
mocou- J
ple
R
S
T
E
B
N
L
U
W/Re5-26
PL-II
Plati-
Pt 100
num
resis-
tance
JPt 100
ther-
mome-
ter
Measurement
Thermocouple:
accuracy
Platinum resistance thermometer:
Setting
There is no difference between the setting accuracy and measurement accuracy.
accuracy
Guaranteed
Same as the setting range.
setting accuracy
range
Thermocouple: ±0.1°C/Ω or ±0.2°F/Ω max.
Influence of
Platinum resistance thermometer: ±1.0°C/Ω or ±1.8°F/Ω max. per line
conductor
resistance
Input shift
–99.9 to 99.9°C/°F (in units of 0.1°C/°F)
Hysteresis
0.0 to 99.9°C/°F for ON/OFF control only (in units of 0.1°C/°F)
Cooling
0.0 to 10.0 (in units of 0.1)
coefficient
Proportional
0.0° to 999.9°C/°F (in units of 0.1°C/°F)
band
Cooling proportional band: Proportional band x Cooling factor
(P constant)
Integral time
0 to 3,999 s (in units of 1 s)
(I constant)
(cannot be set for heating and cooling independently)
Derivative time
0 to 3,999 s (in units of 1 s)
(D constant)
(cannot be set for heating and cooling independently)
Control period
1 to 99 s (in units of 1 s)
Cooling control period: 1 to 99 s (in units of 1 s)
Sampling period 200 ms for 8 control points
1
Celsius
Fahrenheit
–200 to 1,300°C
–300 to 2,300°F
–100 to 850°C
–100 to 1,500°F
0 to 1,700°C
0 to 3,000°F
0 to 1,700°C
0 to 3,000°F
–200 to 400°C
–300 to 700°F
0 to 600°C
0 to 1,100°F
100 to 1,800°C
300 to 3,000°F
0 to 1,300°C
0 to 2,300°F
–100 to 850°C
–100 to 1,500°F
–200 to 400°C
–300 to 700°F
0 to 2,300°C
32 to 4,100°F
0 to 1,300°C
0 to 2,300°F
–100 to 500°C
–100 to 900°F
–100 to 500°C
–100 to 900°F
±0.3% of the process value or ±2°C, whichever is larger ±1 digit max.
±0.3% of the process value or ±3.6°F, whichever is larger ±1 digit max.
±0.3% of the process value or ±0.8°C, whichever is larger ±1 digit max.
±0.3% of the process value or ±1.5°F, whichever is larger ±1 digit max.
Specification
Celsius
–200.0 to 1,300.0°C
–100.0 to 850.0°C
0.0 to 1,700.0°C
0.0 to 1,700.0°C
–200.0 to 400.0°C
0.0 to 600.0°C
100.0 to 1,800.0°C
0.0 to 1,300.0°C
–100.0 to 850.0°C
–200.0 to 400.0°C
0.0 to 2,300.0°C
0.0 to 1,300.0°C
–100.0 to 500.0°C
–100.0 to 500.0°C
Appendix A
0.1
Fahrenheit
(See note 2.)
–300.0 to 2,300.0°F
–100.0 to 1,500.0°F
0.0 to 3,000.0°F
0.0 to 3,000.0°F
–300.0 to 700.0°F
0.0 to 1,100.0°F
300.0 to 3,000.0°F
0.0 to 2,300.0°F
–100.0 to 1,500.0°F
–300.0 to 700.0°F
0.0 to 4,100.0°F
0.0 to 2,300.0°F
–100.0 to 900.0°F
–100.0 to 900.0°F