Replacing the Origio Gas Line Filter
Remove all samples from the chambers before
replacing the filters.
1. To access to the filter compartment,
push the safety lock lever (1) and lift the
upper part of the incubator containing the
chambers.
2. Pull the filter out of the holder and remove
the connectors from each end. Fasten the
connectors to the inlet and outlet of the
Origio Gas Line Filter (2).
3. Ensure the direction of flow shown on
the new Origio Gas Line Filter label (3)
is consistent with the flow direction (4)
shown inside the filter compartment.
4. Place the Origio Gas Line Filter in the filter
holder with the label facing up (5).
5. Lower the upper part of the incubator.
CAUTION: Do not use the unit without a genuine Origio Gas Line Filter.
CAUTION: Insert filter with the filter label facing up. If the filter is not placed
correctly, it can cause excessive gas consumption.
Disposal of the Gas Line Filter
CAUTION: Contamination Hazard
As the filter may have been used for processing and treating infectious substances,
it might be contaminated. The used Gas Line Filter should be placed in a sealed
plastic bag and labeled as biohazard material, then disposed of according to local
requirements.
Section 13
1
2
4
5
3
Service
13
49
Need help?
Do you have a question about the G210 InviCell and is the answer not in the manual?
Questions and answers