Section 3
Safety Warnings
IMMUNITY Test
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
3
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
10
IEC 61326-1
Compliance
Test level
3 A/m
3 A/m
3Vrms, 0.150-
3Vrms, 0.150-
80Hz, 80% AM
80Hz, 80% AM
Modulation, 1s
Modulation, 1s
Dwell Time
Dwell Time
3 V/m
3 V/m
80 MHz to 1000
MHz
3V/m
3V/m
1V/m
1400 MHz to
2000 MHz
1V/m
2000 MHz to
2700 MHz
Electromagnetic environment -
level
Power frequency magnetic fields should
be at levels characteristic of a typical
commercial environment.
Portable and mobile RF communications
equipment should be used no closer to
any part of the G210, including cables,
than the recommended separation
distance calculated from the equation
applicable to the frequency of the
transmitter
Recommended separation distance
d = [ 3,5 ] √P
3
d = [ 3,5 ] √P
3
d = [ 7 ] √P
3
Where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from fixed RF
transmitters, as determined by an
electromagnetic site survey,
less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
guidance
80 MHz to 800 MHz
800 MHz to 2,5 GHz
should be
a