Section 13 - Service; Service Plan; On-Screen Prompts - K-Systems G210 InviCell User Manual

Long-term incubator
Hide thumbs Also See for G210 InviCell:
Table of Contents

Advertisement

Section 13
Service

SECTION 13 - SERVICE

WARNING: DO NOT disassemble or modify any part of the G210 InviCell.
For the reliable and safe operation of this incubator it is strongly recommended that
inspections and services are performed as stated in the Service Plan below. Failing to
follow this plan may cause the unit to stop performing as intended and cause damage to
embryos, blastocysts etc. kept inside the incubator.

Service plan

Conducted by
Component name
Replace Origio Gas Line Filter
Replace HEPA Inline Filter for CO
Replace HEPA Inline Filter for N
Replace O
sensor*
2
Temperature and gas calibration
Replace Pump
Replace CO
sensor*
2
Replace Android battery
*Gas calibration should be performed after replacing O
13
48
User
Every 3
months
X
gas
2
gas
2

On-screen Prompts

1
The service symbol (1) appears in the main
screen when it's time for a service.
The Service menu shows when the unit's
individual parts need to be serviced (2).
This screen (3) also displays hardware and
software versions currently installed and
the unit's serial number.
Authorised Service Representative
Every
Every 3
year
years
X
X
X
X
X
X
and CO
sensors.
2
2
Every 6
years
X