POSITIVE AIRWAY PRESSURE DEVICES
User Guide
English
A

Advertisement

Table of Contents
loading

Summary of Contents for ResMed VPAP ST

  • Page 1 VPAP ™ VPAP ™ AutoSet CS ™ Auto 25 ™ POSITIVE AIRWAY PRESSURE DEVICES User Guide English...
  • Page 2: Adverse Effects

    English S9 VPAP ST indications for use The S9 VPAP ST is intended to provide non-invasive ventilation for patients weighing more than 13 kg with respiratory insufficiency or obstructive sleep apnoea (OSA). The S9 VPAP ST is intended for home and hospital use.
  • Page 3: Control Panel

    • skin rashes. Masks and humidifiers Recommended masks and humidifiers are available on www.resmed.com on the Products page under Service & Support. For information on using your mask or humidifier, refer to the manual supplied with your mask or humidifier.
  • Page 4: Navigating The Menus

    Navigating the menus Refer to illustration C. In general, to navigate the menus: 1. Turn until the parameter you require is displayed in blue. 2. Press . The selection is highlighted in orange. 3. Turn until you see the setting that you require. 4.
  • Page 5: Troubleshooting

    1. Remove the air filter cover from the back of the device. 2. Remove and discard the old air filter. 3. Insert a new ResMed air filter ensuring that it is sitting flat in the air filter cover. 4. Replace the air filter cover.
  • Page 6 SmartStart/Stop is enabled but the device does not stop automatically when you remove your mask Incompatible mask system being used. Only use equipment recommended by ResMed. The patient is using a nasal pillows Disable SmartStart/Stop. mask or pediatric mask with a set...
  • Page 7 Turn the humidifier setting down and empty the water from in accumulation of water in the air the air tubing. tubing. Displays error message: Check ResMed 30/90W Power Supply Unit and fully insert the connector The DC plug is partially inserted into the Fully insert the DC plug.
  • Page 8: Technical Specifications

    Problem/Possible cause Solution The following message is displayed on the LCD after you try to update settings or copy data to the SD card: Card error, please remove SD card and contact service provider SD card is not inserted correctly. Ensure that the SD card is inserted correctly.
  • Page 9 Sound power level 34 dBA with uncertainty of 2 dBA as measured according to EN ISO 17510-1:2009 Nominal dimensions (L x W x H) 153 mm x 140 mm x 86 mm Weight 835 g 90W power supply unit Input range 100–240V, 50–60Hz, Nominal for aircraft use 115V, 400Hz Typical power consumption 70W (80VA) Maximum power consumption 110W (120VA)
  • Page 10 1 m away from the device. Information regarding the electromagnetic emissions and immunity of these ResMed devices can be found on www.resmed.com, on the Products page under Service and Support. Click on the PDF file for your language.
  • Page 11: Pneumatic Flow Path

    If you need information on these disposal systems, please contact your local waste administration. The crossed-bin symbol invites you to use these disposal systems. If you require information on collection and disposal of your ResMed device please contact your ResMed office, local distributor or go to www.resmed.com/environment.
  • Page 12: Limited Warranty

    • Portable diagnostic/screening devices This warranty is only available to the initial consumer. It is not transferable. If the product fails under conditions of normal use, ResMed will repair or replace, at its option, the defective product or any of its components.
  • Page 13 ResMed shall not be responsible for any incidental or consequential damages claimed to have resulted from the sale, installation or use of any ResMed product. Some regions or states do not allow the exclusion or limitation of incidental or consequential damages, so the above limitation may not apply to you.
  • Page 14 If you notice any unexplained changes in the performance of the device, if it is making unusual or harsh sounds, if the device or the power supply are dropped or mishandled, if water is spilled into the enclosure, or if the enclosure is broken, discontinue use and contact your ResMed Service Center.
  • Page 15 • Only use the ResMed 90W or 30W power supply unit. Use the 90W power supply unit to power the system comprising the device, H5i, air tubing, DC/DC converter and battery pack. The 30W power supply unit is designed to power the device only and recommended for traveling.
  • Page 16 ResMed (UK) Ltd 96 Jubilee Ave Milton Park Abingdon Oxfordshire OX14 4RW UK. See www.resmed.com for other ResMed locations worldwide. For patent information, see www.resmed.com/ip. S9, VPAP, AutoSet CS and H5i are trademarks of ResMed Ltd and S9 and VPAP are registered in U.S. Patent and Trademark Office.