Units Without A Service History - ResMed VPAP Auto 25 Service Manual

Hide thumbs Also See for VPAP Auto 25:
Table of Contents

Advertisement

8 Complete the calibrations and tests required when the main PCB has been replaced.
9 Finally, reload the clinical settings to the unit following the procedures from "18 Reload clinical
settings" on page 16.

Units without a service history

In those instances where no session has been previously recorded for a particular unit, or settings
were unable to be downloaded due to a fault with the unit, the default settings will need to be
loaded into the flow generator.
1 Once the new serial number has been loaded into the flow generator, exit the Settings screen.
2 Begin a new recording session for the unit, and complete the calibrations and tests required
when a main PCB is replaced.
3 Once the testing is complete, stop the recording.
4 Select Flow Generator Settings... in the XPlorer bar, or click on the Flow Generator Setting
Dialog icon
5 Select the appropriate product code for the unit from the drop down menu, then click on Read
Default Settings.
Read default
settings
6 The screen will then be populated with the units default settings.
7 Click on Reload to flow generator to load them into the unit.
8 Exit the Settings screen.
9 Once again, press the Record button and ensure the most recent session (in which the clinical
settings and results were recorded) is highlighted, then click on the Select button
10 Click on Reload settings at the bottom of the Xplorer bar.
11 The following window will be displayed:
24
VPAP™ Service Manual
at the top of the screen.
.