16 Finalise manual check results

In the XPlorer bar at the left of the screen, all tests that are completed without the service software
are listed under the title "Manual Check".
1 Click on the name of each manual check. A screen such as that shown below will be displayed:
For each manual check, indicate whether the unit passed or failed.

17 Set clock

Note:
Ensure that the PC used to service the flow generator is set to the correct date and time
prior to pressing the Set time and date icon.
1 Select Set time and date... in the XPlorer bar, or click on the Read and adjust flow generator
clock icon
.
2 The following window will appear:
3 General inspection and servicing