Pressure Calibration Test - ResMed VPAP Auto 25 Service Manual

Hide thumbs Also See for VPAP Auto 25:
Table of Contents

Advertisement

3 Wait for the prompt to remove the card.
4 Remove the data card and confirm that the unit returns to the Standby screen.
5 Turn off the unit and disconnect it from the mains power.
6 Remove the data card module.
9

Pressure calibration test

Caution
Make sure the flow generator has been warmed up for at least 15 minutes before
proceeding with the remaining tests.
!
1 With the unit still switched off, remove the data card module and refit the ResScan USB
adaptor, with USB cable attached.
2 Turn on the unit and allow it to complete the self test.
3 Return to the service software and press the Record button
4 The following screen will then appear, indicating the existing service records for the flow
generator.
5 The most recently created record (the flow generator settings recorded earlier) will be
highlighted.
6 Click on the Select button
7 The "Select service agent screen", as shown on page 9, will appear once again.
8 Select the name of the service agent and enter the SR number.
9 All calibration and tests, including tests completed without service software are then listed in
the Xplorer bar.
10 Select Pressure calibration test in the XPlorer bar, or click on the Perform the pressure
calibration test icon
Note:
At the top of the screen at the beginning of each procedure is an alert reminding you to
reload the clinical settings to the unit after completing the necessary test and calibration
procedure. Click the OK button once you have read the alert.
11 Set up the test equipment as instructed by the test software.
12 Select Next and follow the steps as instructed.
13 If the flow generator fails this test, repeat the pressure calibration, ensuring the test set-up is
correct.
to continue recording the service results to the same record.
.
.
3 General inspection and servicing