5 Set up with the display and adjustment module
Abb. 26: Mounting of the display and adjustment
module
5.3

Adjustment system

1
Abb. 27: Display and adjustment elements
1
LC display
2
Indication of the menu item number
3
Adjustment keys
Key functions
[OK] key:
– Move to the menu overview
– Confirm selected menu
– Edit parameter
– Save value
[->] key to select:
– Menu change
– Select list entry
– Select editing position
[+] key:
– Change value of the parameter
[ESC] key:
– Interrupt input
– Return to superordinate menu
26
2
1.1
3
Adjustment system
The sensor is adjusted via the four keys of
the display and adjustment module. The LC
display indicates the individual menu items. The
functions of the individual keys are shown in
the above illustration. Approx. 10 minutes after
the last pressing of a key, an automatic reset to
measured value indication is triggered. Any va-
lues not confirmed with [OK] will not be saved.
5.4

Setup steps

Parameter adjustment
To carry out this adjustment, the distance is
entered for full and empty vessel. If these values
are not known, an adjustment with distances,
for example, of 10 % and 90 % is also possible.
Starting point for these distance specifications
is the lower edge of the flange (flange versi-
ons), for all other versions the lower edge of the
transducer.
The actual product level during this adjustment
is not important, because the min./max. adjust-
ment is always carried out without changing
the product level. These settings can be made
ahead of time without the instrument having to
be installed.
In the main menu item "Basic adjustment", the
individual submenu items should be selected
one after the other and provided with the correct
parameter values.
Carry out min. adjustment
Proceed as follows:
1. Move from the measured value display to
the main menu by pushing [OK].
Basic adjustment
Display
Diagnostics
Service
Info
2. Select the menu item "Basic adjustment"
with [->] and confirm with [OK]. Now the
menu item "Min. adjustment" is displayed.
Min. adjustment
0.00 %
=
5,000 m(d)
4,000 m(d)
VEGASON S 62 •