Resolution, general
Ultrasonic frequency
Interval
Abstrahlwinkel at -3 dB
Adjustment time
11)
Accuracy
10 mm
4 mm
1 m
-4 mm
-10 mm
Abb. 29: Accuracy VEGASON S 62
Ambient conditions
Ambient, storage and transport tempe-
rature
Process conditions
Vessel pressure
Process temperature (transducer tem-
perature)
Vibration resistance
Electromechanical data
Cable entry
Spring-loaded terminals
Display and adjustment module
Power supply and data transmission
Indication
Adjustment elements
Protection rating
Ʋ unassembled
Ʋ Mounted into the sensor without cover IP40
Materials
Ʋ Housing
Ʋ Inspection window
Time to output the correct level (with max. 10 % deviation) after a sudden level change.
11)
Tested according to the guidelines of German Lloyd, GL directive 2.
12)
VEGASON S 62 •
2 m
3 m
4 m
5 m
max. 1 mm
70 kHz
> 2 s (dependent on the parameter adjustment)
11°
> 3 s (dependent on the parameter adjustment)
better than 0.2 % or ±4 mm (see diagram)
-40 ... +80 °C
-0.2 ... 2 bar/-20 ... 200 kPa
-40 ... +80 °C
mechanical vibrations with 4 g and 5 ... 100 Hz
1 x cable gland M20 x 1.5 (cable: ø 5 ... 9 mm), 1 x blind
plug M20 x 1.5
for wire cross-section up to 2.5 mm²
through the sensor via sliding contacts
LC display in dot matrix
4 keys
IP20
ABS
Polyester foil