The batteries
Two 12V 2,6 Ah maintenance-free lead-acid batteries are used. They have an expected life span of
three years.
If the output voltage of one of the batteries is at least 5V less than that of the other, it can be
assumed the battery is faulty. If the voltage between the batteries varies, there is something wrong
with them.
Guarantee
No guarantee is provided for the batteries. However, we do ask that you report any complaints to us,
stating the identifi cation number on the battery.
Code "scroll"
F A U L T
ion
Display
ol
Toggle delay: self test lift
F A U L T
If the lift is doing the selftest each time that the toggle is used and the lift does not move,
the batteries are not connected.
Fuse
A fuse of 20 amp is installed in the "-" line.
Code "FuSE"
Display
Fuse blown or battery disconnected
F A U L T
The fuse is blown or the battery is disconnected.
When you keep activating the control, the display will start scrolling indicating that it is doing a self
test (see page 8).