Charging / Power Supply - Handicare 1100 Quick Reference Manual

Hide thumbs Also See for 1100:
Table of Contents

Advertisement

Handicare 1100
11
|

Charging / Power supply

The battery charger is suitable for an input voltage of between 100V and 240V.
The output voltage is 33 V=. The charging current is 1 A.
Code "1" and "2" and "7"
Display
Requires charge
F A U L T
Charging
There are diff erent diagnostic codes if the lift requires charge.
Code 1 and 7:
When the battery voltage is very low (maybe down to 10% capacity left in batteries), this fault code
will appear.
When the lift docks on to charge contacts, the lift will not drive off the contacts.
1 will fl ash.
Lift will only drive off if battery voltage has recovered approximately three quarter capacity.
Another way of driving off the contacts is if lift is powered "off " then back "on" but as soon as lift
docks on contacts, process will repeat again.
This fault code is linked to fault code 7 but the code 7 is not usually seen as code 1 takes priority.
Code 2:
The lift is off charge and gives an audible signal.
26
Action
Display
Action
!
Off charge
F A U L T
D
Display
Action
30 min.
Low battery voltage
F A U L T
D

Advertisement

Table of Contents
loading