6
|
Chair position switch
Code "-"
F A U L T
Display
Action
CLICK!
Chair in swivel position
F A U L T
The lift does not move but the automatic swivel functions
Powered swivel:
•
When moving the toggle in up direction, the chair swivels: The display shows code "8." and after 8
seconds its shows code "4", when the lift is not at an endpoint.
•
When moving the toggle in down direction, the chair swivels: The display shows code "9." and
after 8 seconds its shows code "6". when the lift is not at an endpoint.
Manual swivel:
•
When moving the toggle in up direction, the display shows code "8." and the lift is not moving.
•
When moving the toggle in down direction, the display shows code "9." and the lift is not moving.
Symptoms
Lift moves up
Lift moves down
Powered swivel
Powered footrest
16
D
Connection
Standard
A1-4
4,3 V
A1-5
4,3 V
A1-6
4,3 V
A1-10
4,3 V
Function
Cause
The chair position switch (S14) is defect/off
•
no
•
Wire break between A9-1, S14,A9-2, A8-5 and A15
no
yes
yes
Faulty
4,3 V
0 V
0 V
4,3 V