Handicare 1100
4
|
Toggle
The lift is operated by the toggle on the armrest. The toggle as well as the key switch and the button
for the powered footrest can be swapped from right to left on side. The lift will not ride when the
armrests are in the folded position. It does however operate on the remotes. During the ride the
diagnostic display show "8" or "9".
An installation toggle loom can be used to ride the power Unit without chair.
Code "-"
F A U L T
Display
Chair in swivel position
F A U L T
Symptoms
Lift moves to the right
Lift moves to the left
Powered swivel
Powered footrest
Symptoms
Lift moves to the right
Lift moves to the left
Powered swivel
Powered footrest
Symptoms
Lift moves to the right
Lift moves to the left
Powered swivel
Powered footrest
10
Action
D
CLICK!
Connection
A1-3
A1-8
A1-2
The main power of the toggle is on A1-8 (11,6 V=).
Function
Cause
•
The toggle is defect
no
•
The armrest switch S10 or S11 is defect
no
•
A1-8, A8-8, A20-2, S11, A20-1, A24-1, S10, A24-2 and
toggle (S9)
no
•
The armrest is folded
yes
Function
Cause
•
Wire break between A1-3, A8-3 and toggle (S9)
no
•
The toggle is defect
yes
one direction
yes
Cause
Function
•
Wire break between A1-2, A8-2 and toggle (S9)
yes
•
The toggle is defect
no
one direction
yes
Standard
Toggle right
0 V
11,6 V
11,6 V
11,6 V
0 V
0 V
Toggle left
0 V
11,6 V
11,6 V