Handicare 1100
5
|
General safety line
The next components are part of the general safety line:
•
The keyswitch
•
The emergency stop
•
The chair position switch
When the keyswitch and the emergency stop are activated, the lift won't move in both directions and
the powered swivel and the powered footrest do not function.
Code "0" the lift is not moving and the powered options don't function
F A U L T
ion
Display
Key switch off
F A U L T
When the toggle is pushed in downwards and upwards direction, the lift does not move and the
powered swivel and the powered footrest don't function, the display shows code "0".
Symptoms
Lift moves up
Lift moves down
Powered swivel
Powered footrest
12
Action
Display
Action
ON
Emergency stop activated
F A U L T
Function
Cause
The keyswitch (S12) is defect / off
•
no
The emergency stop (S13) is defect / off
•
no
•
Wire break between A9-1, A27-1, S12, A27-2, A11-1,
S13, A11-3, A8-10 and A1-10
no
no
Connection
Standard
A1-4
4,3 V
A1-5
4,3 V
A1-6
4,3 V
A1-10
4,3 V
Faulty
10,7 V
10,7 V
10,7 V
0 V