Total Trunk: Service & Maintenance Manual - rev. 1.5

11. FUNCTIONAL TESTS

Every time maintenance and/or repairs are performed on a machine, it is necessary to carry out a
series of mechanical and electrical tests to ensure:
• that the medical device is working properly;
• that the medical device is working safely;
• that the exercises are performed according to the biomedical specifications for which the
medical device was designed.
11.1.

ELECTRICAL SAFETY TEST

Technogym ensures its own safety standards by means of the following tools and activities:
• ISO 9001/2000 certification of the company;
• certification of the device in compliance with medical regulations 60601-1 and 60601-1-2;
• electrical testing of each device manufactured, according to the provisions of the
aforementioned standards;
• training of technical staff who perform technical service;
• use of original Technogym spare parts for all service operations;
Based on the risk analysis carried out by the After Sale department regarding installation and
technical service, it is still necessary to use a tester to perform a safety test on the grounding ring.
This test must ensure a continuous earth connection (resistance less than 4 Ohm) between the
grounding pole of the mains lead (connected to the machine) and the 6 points listed below:
1. screw fastening the guard to the frame;
2. screw fastening the loading cell to the exercise lever;
3. shoulder roller holding screw;
4. seat adjustment knob holding screw;
5. thigh lock stop screw;
6. mobile PC arm holding screw.
For more complete explanations, see the boxes in the figure below.
Pagina 11.1