Process Input - Omega DP6000 User Manual

Dual-line process meter
Table of Contents

Advertisement

DP6000 Process Meter
NORMAL MODE
REJECTION
ISOLATION
OVERVOLTAGE
CATEGORY
ENVIRONMENTAL Operating temperature range: -40 to 65°C
CONNECTIONS
ENCLOSURE
MOUNTING
TIGHTENING
TORQUE
OVERALL
DIMENSIONS
WEIGHT
WARRANTY

Process Input

INPUTS
ACCURACY
TEMPERATURE
DRIFT
SIGNAL INPUT
CONDITIONING
MULTI-POINT
LINEARIZATION
PROGRAMMA-
BLE EXPONENT
ROUND H TANK
Greater than 60 dB at 50/60 Hz
4 kV input/output-to-power line
500 V input-to-output or output-to-P+ supply
Installation Overvoltage Category II:
Local level with smaller transient overvoltages than
Installation Overvoltage Category III.
Storage temperature range: -40 to 85°C
Relative humidity: 0 to 90% non-condensing
Removable screw terminal blocks accept 12 to 22 AWG
wire, RJ45 for external relays, digital I/O, and serial
communication adapters.
1/8 DIN, high impact plastic, UL 94V-0, color: black
1/8 DIN panel cutout required:
3.622" x 1.772" (92 mm x 45 mm)
Two panel mounting bracket assemblies are provided.
Screw terminal connectors: 5 lb-in (0.56 Nm)
4.68" x 2.45" x 5.64" (119 mm x 62 mm x 143 mm)
(W x H x D)
9.5 oz (269 g)
3 years parts & labor
Field selectable:
0-20, 4-20 mA, 10 V (0-5, 1-5, 0-10 V), Modbus PV (Slave)
±0.03% of calibrated span ±1 count,
square root & programmable exponent accuracy
range: 10-100% of calibrated span
0.005% of calibrated span/C max from 0 to 65C ambient,
0.01% of calibrated span/C max from -40 to 0C ambient
Linear, square root, programmable exponent, or
round horizontal tank volume calculation
2 to 32 points for PV or PV1
2 to 8 points for PV2 (Dual-scale Level feature)
1.0001 to 2.9999
Diameter & Length: 999.999 inch or cm calculates volume
in gallons or liters respectively.
11