Introduction - Omega DP6000 User Manual

Dual-line process meter
Table of Contents

Advertisement

DP6000 Process Meter
Table of Figures
Figure 1. 1/8 DIN Panel Cutout Dimensions ..................................... 17
Figure 2. Panel Mounting Details ...................................................... 17
Figure 3. Meter Dimensions - Side View ........................................... 18
Figure 4. Meter Dimensions - Top View ............................................ 18
Figure 6. Transmitter Supply Voltage Selection .............................. 20
Figure 8. Power Connections ............................................................ 21
Figure 11. Voltage Input Connections .............................................. 23
Figure 12. Relay Connections ........................................................... 23
Figure 13. AC and DC Loads Protection ........................................... 24
Figure 14. Low Voltage DC Loads Protection .................................. 24
Figure 15. F4 Digital Input Connections ........................................... 25
Figure 16. 4-20 mA Output Connections .......................................... 25
Figure 18. External Relays Module Connections ............................. 26
Figure 19. Digital I/O Module Connections ....................................... 27
Figure 20. Interlock Connections ...................................................... 28

INTRODUCTION

The DP6000 is a multi-purpose, easy to use digital process meter ideal for level, flow
rate, temperature, or pressure transmitter applications. It accepts current and voltage
signals (e.g. 4-20 mA, 0-10 V). Three of the front panel buttons can be custom-
programmed for a specific operation. The analog input can be scaled to display the
process in two different scales; for example the main display could indicate level in
feet and the second display could indicate the volume in gallons.
The basic model includes an isolated 24 VDC transmitter power supply that can be
used to power the input transmitter or other devices. An additional isolated 24 VDC
power supply is included with the 4-20 mA output option. A digital input is standard.
A fully loaded DP6000 meter has the following: four SPDT relays, 4-20 mA output, and
two 24 VDC power supplies. The DP6000 capabilities may be enhanced by adding the
following external expansion modules: four SPST relays (creating an eight-relay process
meter), two digital I/O modules with four inputs and four outputs each, and USB, RS-232
or RS-485 communication adapters.
The eight relays can be used for alarm indication or process control applications, such as
pump alternation control. The 4-20 mA isolated output, Modbus RTU serial
communications, and digital I/O options make the DP6000 an excellent addition to any
system.
8
Instruction Manual
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 
 

Advertisement

Table of Contents
loading

Table of Contents