General Information; Specifications; Analog Input - Omega DP301AN User Manual

Voltmeter
Table of Contents

Advertisement

1.0

GENERAL INFORMATION

The two versions of voltmeters are different pinout versions of the same 3 1/2
digit panel meter for applications which require a compact, quality dc voltmeter.
Standard features include HOLD function, selectable decimal point location and
LAMP TEST.
Each model contains the circuitry for all ranges. The range is changed by
opening and closing the appropriate switches on the printed circuit board via
installation of push-on jumpers.
1.1
MODEL DP301AN
Model DP301AN is pin compatible with newer voltmeters DP302, DP3001 and
DP3002 series.
Models DP301AN has a resolution of 1 part in ±1999 counts, or 0.05% of full
scale. By using an external dc voltage reference, any of these ranges can be
configured for 3 wire ratio measurement with readout from 0 to 1.999.
1.2
MODEL DP301AN-AC
Model DP301AN-AC, for AC power line monitoring, has a screw terminal barrier
strip for signal and power. This model uses a half wave average converter
calibrated to read RMS. It offers input ranges of 199.9 and 650 V ac, with
resolution of 100 mV and 1 V, respectively. Maximum rated input in either range is
650 V RMS or 920 Vp.
2.0

SPECIFICATIONS

2.1

ANALOG INPUT

Model "AN"
Signal ranges
Bias current
Maximum input
Resolution
Input resistance
Configuration
Zero
Span adjustment
Model "AN-AC"
Signal ranges
Bias current
Overvoltage protection
Resolution
Input resistance
Configuration
Zero
Frequency
±199.9mV
±1.999V
50 pA
50 pA
130 Vp
130 Vp
100 uV
1 mV
1 GΩ
1 GΩ
Differential, bipolar
Automatic
±5%
199.9 V ac
650 V ac
1 pA
1 pA
650 V ac
650 V ac
100 mV
1 V ac
1 MΩ
1 MΩ
Half wave average, RMS calibrated
Automatic
47 to 1000 Hz
1
±19.99V
±199.9V
5 pA
1 pA
250 Vp
250 Vp
10 mV
100 mV
1.1 MΩ
1.0 M^N8,50

Advertisement

Table of Contents
loading

Table of Contents