Bits:
O:
2:
3:
Set when a screen dump (in REMOTE only) is finished
Set when calibration is done (in REMOTE only)
Average END
ERROR Status
Byte (STB 6):
This status byte contains the coded value of the most recently detected
error. The individual bits have no well defined meaning. Whenever MASK
6 is non-zero, and the code in STB 6 is set to a non-zero value, the
ERROR bit (DIO-6) in STBI is set.
Code:
0:
I:
No error
Not all 9400A responses have been read upon receipt of a
new command string, terminated with <END>. Output buffer
has been flushed.
Syntax errors:
60:
I00:
i0: Invalid separator or too many parameters
II: Invalid header
12: Invalid number format
13: Invalid keyword
14: Invalid block
15:
2 or more strings in the same command
20:
Command permission error; REMOTE only command has been
received while the 9400A is in LOCAL, or SCREEN control
command has been received while SCREEN control is LOCAL
30:
The 9400A has received a command for an option that is
not installed
40:
Semantic
error: false number of parameters
or false
parameter in a command
50:
Environment
error: the 9400A is not set to the proper
status to accept the received command (trace OFF, false
format...)
Descriptor error; an inconsistency has been detected in
the data received with a WRITE descriptor command
Command not yet implemented
i)
STB
, [ I to
6 ]
?
Reads the contents of one of the 6 status bytes and clears it. STBI
is the Main Status Byte. STB 4 is not cleared.
or
STB
,
?
Reads all 6 status bytes in the order I to 6 and clears them (except
STB 4).