READ (RD)
, < CHANNEL 1.DATA
< CHANNEL 2.DATA
< MEMORY C.DATA
< MEMORY-D.DATA
(CI.DA) > , <Parameter list>
(C2.DA) >
(MC.DA)
(MD.DA)
<Parameter list> = <intval> , <# values> , <addr> , <sweep #>
transfer the data values of the indicated waveform from the 9400A to
the host computer. An explanation of the optional parameter list is
given below.
or
READ (RD)
, < CHANNEL 1.TIME
< CHANNEL-2.TIME
< MEMORY C.TIME
< MEMORY-D.TIME
(CI.TI)
(C2.TI)
(MC.TI)
(MD.TI)
transfer
the trigger time(s) of the indicated
waveform
from the
9400A to the host computer. See Section 7.8 for the format of this
data block.
or
READ (RD)
, < CHANNEL I.* (CI.*) > , <Parameter list>
< CHANNEL-2.* (C2.*)
< MEMORY C.* (MC.*)
< MEMORY-D.* (MD.*)
transfer ALL visible data of the indicated waveform from the 9400A
to the host computer. Data are transferred in the order descriptor,
data, time(s).
<Parameter list> = <intval> , <# values> , <addr> , <sweep #>
<intval>
= [ 1 to 16000 ]
Interval between data points to be
read, for example:
I = read all points
4 = leave out 3 of 4 data values
<# values>= [ 0 to 32000 ]
Number of data values to read
<addr>
= [ -32000 to 32000 ]
Address
of
first
data
relative to the left hand
the screen
point
side of
<sweep #> = [ 0 to 250 ]
Sweep number in SEQNCE
numbered from
1 to max. # sweeps
0 = read all sweeps
waveforms,