I data-128
°fret
2°oil
20° 1
V = gain
32
25
vgain + 80
- The first expression "(data - 128)/32" transforms the unsigned data
value to a signed value, in units of vertical divisions.
- The second expression
"(offset - 200)/25" translates
the internal
coding of the offset into a signed offset, units: vertical divisions.
- The third expression
"200/(vgain
+ 80)" takes into account
the
variable gain. It reduces, to the value 1 when "vgain" assumes the
"calibrated" value of 120.
Examples:
gain = I00 mV byt e 0 of descriptor = 26
offset = 200
no offset
vgain = 120
calibrated gain
The formula reduces to
V = I00 mV (data - 128)/32
data
volts
0
-400 mV
128
0 mV
129
3.12 mV
160
+I00 mV
240
+350 mV
255
+396.88 mV
gain = 500 mV byte 0 of descriptor = 28
offset = 250
offset
= + 2 div
vgain = 20
variable gain
= .500 (equivalent to
MULTIPLYING the
vertical scale by 2)
The formula reduces to
V = 500 mV ((data - 128)/32 - 2) *
data
volts
0
-6.0 V
64
-4.0 V
128
-2.0 V
192
0 V
193
31.25 mV
200
.25 V
240
I. 5 V
255
i. 969 V
7-51
Remote Operations