LeCroy 9400A Manual page 53

Digital oscilloscope
Table of Contents

Advertisement

Number of Segments
Points/Seg~nent
8
2501
15
2001
31
i001
62
497
125
241
250
I01
SEQUENCE TRIGGER MODE
NUMBER OF SEGMENTS VS. RECORD LENGTH (TIME BASE: 20 usec)
Table 5.2
Neither the CHAN 1 nor CHAN 2 display is updated when the 9400A is
in the SINGLE or SEQNCE trigger mode, i.e. when no further data
are acquired.
Vertical
positioning
of the displayed
trace may
nevertheless
be modified
via the OFFSET control (32). The VAR
vernier (28) also remains
active. However,
no other parameter
modifications, such as vertical sensitivity or time changes, will
alter the display of a currently acquired waveform in CHAN 1 or
CHAN 2.
Of course, all parameters
may be modified
during this time by
manipulating
the appropriate
front
panel controls,
but such
modification
- indicated
by parameter
changes
in the Abridged
Front Panel Status field (IV) - will only be used when acquiring
the next trace.
Whenever the 9400A is in the NORM or AUTO trigger mode, data are
continuously
acquired
and the display
rapidly
updated.
All
modifications in acquisition parameters are thus followed quickly
by subsequent
waveform
acquisition
which
results
in their
appearing to the user as changes to the CHAN 1 or CHAN 2 display.
TRIG'D and READY LEDs (31) - The TRIG'D LED indicates
Whenever
the
digitizing has stopped (normally after a valid trigger). The READY LED
indicates
that the trigger circuit has been armed and the 9400A is
currently
digitizing
input signals. Upon receiving
a valid trigger
signal, it will continue digitization until the trigger conditions have
been satisfied and will then display the acquired waveform.
DELAY (34) - Adjusts the degree of pre- or post-trigger
delay when
recording signals in the acquisition memories. Delay operation is via a
single continuously
rotating
knob. Turning this knob slowly allows
minute adjustment of the trigger point; turning it quickly results in
rapid trigger point movement. The DELAY control enables pre-trigger
adjustment,
displayed
in %. Pre-trigger
adjustment
up to 100% full
scale, and post-trigger
adjustment
up to i0,000 divisions
in 0.02
division increments are available.
Manual Operation
5-11

Advertisement

Table of Contents
loading

Table of Contents