The 9400A sets the VALUE ADAPTED bit:
- if a positive out-of-range value is given.
- if a negative value corresponding to more than i0000 divisions is
given.
4) TRIG LEVEL (TRL) , < ?
, < [ -5.00 DIV to 5.00 DIV ] >
when
the oscilloscope
is set
to one of the
internal
trigger
sources
(CHANNEL_I or
CHANNEL_2).
, < [ -2.00 V to 2.00 V ]
>
when
the oscilloscope
is set to in the EXT
trigger source.
, < [ -20.0 V to 20.0 V ]
>
when
the oscilloscope
is set to the
EXT/10
trigger source.
When the oscilloscope
is set to LINE trigger source, this command
has no meaning and no error will be reported. In the case of POS_NEG
triggering, only a positive value is meaningful.
The 9400A sets the VALUE ADAPTED bit:
- if an out-of-range value is given.
The 9400A sets the ENVIRONMENT ERROR:
- if the DIV suffix is sent instead of V or if the V suffix is sent
instead of DIV.
5) TRIG COUPLING (TRC)
, < ?
>
< AC
>
*
< DC
>
*
< LF REJ (LF)
*
< HF-REJ
(HF)
*
6) TRIG MODE (TRM)
, < ?
>
< SEQNCE (SE)
>
*
< AUTO (AU)
>
*
< NORM (NO)
>
*
< SINGLE (Sl)
>
*
The 9400A sets the ENVIRONMENT ERROR:
- if SEQNCE is sent while the 9400A is in Interleaved Sampling.
7) TRIG SOURCE
(TRS)
, < ?
< CHANNEL i (CI)
*
< CHANNEL-2 (C2)
*
< LINE (eI)
>
*
< EXT (EX)
>
*
< EXT/IO (E/IO)
>
*
Remote Operations