The 9400A sets the VALUE ADAPTED bit:
- if an out-of-range value is given.
The 9400A sets the ENVIRONMENT ERROR:
- If the SELECTed trace is OFF.
6) TIME MAGNIFIER
(TM) , < ?
< [ 0 to6 ] >
TIME MAGNIFIER
is applied to the SELECTed trace. It may only be
applTed to traces EXPAND A or EXPAND B.
The value
0 corresponds
to no expansion.
Each increment
of I
corresponds
to the next lower time base value (relative
to the
original trace). The value 6 therefore
allows an expansion
by
factor of I00.
The 9400A sets the VALUE ADAPTED bit:
- if an out-of-range
positive
value
generate a semantic error.
is given.
Negative values
The 9400A sets the ENVIRONMENT ERROR:
- if the SELECTed trace is OFF.
- if the SELECTed trace is neither EXPAND A nor EXPAND B.
7) HOR POSITION (HP) , <
>
, < [ 0.0000 DIV to i0.0000 DIV ] >
when the source trace corresponds
to a single
waveform
(acquired
in single
shot
or with
INTERLEAVED ON).
, < [ 1 to max ] >
when the source trace corresponds
to multiple
waveforms (acquired in SEQNCE).
The parameter
indicates
the number
of the
sequence to be displayed. "max" depends on the
selected number of segments (see SEGMENTS).
HOR POSITION will be applied to the SELECTed trace. It may only be
applied to traces EXPAND A or EXPAND B. The parameter in the range
0 to I0 divisions
corresponds
to The CENTER of the intensified
region
on the original
trace.
The smallest
possible
step
is
0.0004 div.
The 9400A sets the VALUE ADAPTED bit:
- if an out-of-range value is given (Interleaved Sampling or Single
Shot).
- if a positive out-of-range value is given (SEQNCE). Negative
zero values generate a semantic error.
7-18
Remote Operations