LeCroy 9400A Manual page 176

Digital oscilloscope
Table of Contents

Advertisement

<s-type>
=<I>
<3>
<5>
<7>
<9>
Smoothing
type
Default:
3-pt
smoothing
Any other number generates a semantic error.
<m-fact> = [ 0.01 to 9.99 ]
Multiplication factor
Default: 1.00
Any value
smaller
than
0.01
or
greater
than
9.99
is
adapted
to
the
nearest
legal
value.
<a-const> = [ -9.99 to 9.99 ]
Additive constant
Default: 0.00
Any value
smaller
than
0.01
or
greater
than
9.99
is
adapted
to
the
nearest
legal
value.
NOTE: <a-const> is interpreted as the number of vertical divisions
of <sourcel>
Whenever
a parameter
is not specified,
the default
value
is
substituted.
Examples:
RDF,AVG,SUMMED,,C2
Configures the SELECTed trace (which must be
FUNCTION
E or FUNCTION
F) for the summed
averaging of CHANNEL 2 with a maximum of 1250
data points
over IO00 sweeps.
By default,
overflow rejection and dithering are off.
RDF,ARI,PRD,2500,CI,MD,2.00,1.00
Configures the SELECTed trace to compute the
waveform 2.00*(CHANNEL 1 + 1 div)*Memory_D.
maximum of 2500 of data points (equidistantly
distributed over the screen) are to be used.
REDEFINE,FNC,INTEG,25000,FE
Configures the SELECTed trace (which must be
FUNCTION F, since FUNCTION E is specified as
source waveform) to compute the integral over
the waveform in FUNCTION_E, using a maximum of
25000 data points. Note that the use of 25000
data points implies that all data points on
the screen are to be used, regardless of the
time-base setting.
10-12
WPOI Waveform Processing Option