Alarm Relays: The nine alarm-circuit connector pins connect to the
internal alarm relay contacts. Each set of three pins provides one set of
Form C relay contacts. Each relay has both normally open and normally
closed contact connections. The contact connections are shown in Table
3-2. They are capable of switching up to 3 amperes at 250 VAC into a
resistive load. The connectors are:
Threshold Alarm 1:
Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is
below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
Threshold Alarm 2:
Can be configured as high (actuates when concentration is
above threshold), or low (actuates when concentration is
below threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).

Table 3-2: Alarm Relay Contact Pins

Pin
Contact
45
Threshold Alarm 1, normally closed contact
28
Threshold Alarm 1, moving contact
46
Threshold Alarm 1, normally open contact
42
Threshold Alarm 2, normally closed contact
44
Threshold Alarm 2, moving contact
43
Threshold Alarm 2, normally open contact
36
System Alarm, normally closed contact
20
System Alarm, moving contact
37
System Alarm, normally open contact