2. Connect four BNC(f)-BNC(f) Adapters directly to another end of the 16048A.
3. Connect the specied termination or standard resistor directly to the 4263B UNKNOWN
Connectors.
4. Repeat step 3 as instructed.
Impedance Measurement 2m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when the 2m test leads are used.
Note
Perform the Impedance Measurement 0m Adjustment prior to performing this
adjustment.
Perform the same setup procedures for the 16048D as used for the 16048A.
Impedance Measurement 4m Adjustment
This adjustment updates the data to adjust the measurement impedance to the standard values,
when the 4m test leads are used.
Note
Perform the Impedance Measurement 0m Adjustment prior to performing this
adjustment.
Perform the same setup procedures for the 16048E as used for the 16048A.
2-10
Adjustments