Transducer Section

The transducer section measures the vector voltage across the DUT, and the current owing
through the DUT by converting the current to voltage. The transducer also includes a switch
used by the Option 001 (Transformer Parameter Measurement), and the contact check voltage
source.
Figure 4-5 shows the transducer block diagram.
The voltage across the DUT is measured at the H
and L
connectors using a di erential
POT
POT
ampli er.
The current owing through the DUT and owing into the L
connector is converted to
CUR
voltage by an I-V converter. The di erential ampli er measures the voltage between an
I-V converter output and the inverting input of the converter. The inverting input of the
converter, thus L
, is at virtual ground, because of the very high input impedance and very
CUR
high open loop gain of the operational ampli er used in the I-V converter.