Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The POC-624-02 series device is intended for use in the electromagnetic environment
specified below. The customer or the user of the POC-624-02 series device should ensure
that it is used in such an environment.
Immunity
test
Conducted
RF
IEC 61000-4-
6
Radiated RF
IEC 61000-4-
3
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption and reflection from structures, objects and people.
POC-624-02 Series User Manual
IEC 60601 test
Compliance
level
level
3 and 6 Vrms
3 and 6 Vrms
150 kHz to 80
MHz and specific
150 kHz to 80
ISM, AM fre-
MHz and spe-
quency
cific ISM, AM
3 and 6 Vrms
frequency
3V/m
3V/m
80 MHz to 2.7
80 MHz to 2.7
GHz, and
GHz, and specific
specific fre-
frequency
quency
Electromagnetic environment –
guidance
Portable and mobile RF communi-
cations equipment should be used
no closer to any part of the POC-
624-02 series device, including
cables, than the recommended sep-
aration distance calculated from the
equation applicable to the fre-
quency of the transmitter.
Recommended separation dis-
tance
d = 1,2
P
d = 1,2
80 MHz to 800 MHz
P
d = 2,3
800 MHz to 2,5 GHz
P
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmit-
ter manufacturer and d is the rec-
ommended separation distance in
metres (m).
Field strengths from fixed RF trans-
mitters, as determined by an elec-
tromagnetic site survey,
less than the compliance level in
each frequency range.
Interference may occur in the vicin-