Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The POC-624-02 series device is intended for use in the electromagnetic environment
specified below. The customer or the user of the POC-621-02 series device should ensure
that it is used in such an environment.
IEC 60601 test
Immunity test
level
Electrostatic dis-
±8 kV contact
charge (ESD)
±15 kV air
IEC 61000-4-2
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power
supply lines
±1 kV for input/out-
put lines
±1 kV line(s) to
line(s)
Surge
±2 kV line(s) to
IEC 61000-4-5
earth
Interruptions and
voltage varia-
Voltage Dips:
tions on power
i) 0% reduction
supply input lines
for 0.5 cycle at
IEC 61000-4-11
50Hz,
0% reduction
for 1 cycle at 50Hz
Performance
Criterion B
ii) 70% reduction
for 25/30 cycles at
50/60Hz
Performance
Criterion C
Power frequency
Voltage Interrup-
(50/60 Hz)
tions:
magnetic field
0% reduction
IEC 61000-4-8
for 250/300 cycles
at 50/60Hz
Performance
Criterion C
30A/m
Compliance
Electromagnetic environ-
level
ment –guidance
±8 kV contact
Floors should be wood, con-
±15 kV air
crete or ceramic tile. If floors
are covered with synthetic
material, the relative humidity
should be at least 30%.
±2 kV for
Main power quality should be
power supply
that of a typical commercial or
lines
hospital environment.
±1 kV for input/
output lines
If the user of the POC-624-02
±1 kV line(s) to
series device requires contin-
line(s)
ued operation during main
±2 kV line(s) to
power interruption, it is rec-
earth
ommended that the POC-
621-02 series device be pow-
Voltage Dips:
ered from an uninterruptible
i) 0% reduc-
power supply.
tion for 0.5
cycle at 50Hz,
Power frequency magnetic
0% reduc-
fields should be at levels
tion for 1 cycle
characteristic of a typical
at 50Hz
location in a typical commer-
cial or hospital environment.
Performance
Criterion B
ii) 70%
reduction for
25/30 cycles at
50/60Hz
Performance
Criterion C
Voltage Inter-
ruptions:
0% reduc-
tion for 250/
300 cycles at
50/60Hz
Performance
Criterion C
30A/m