T-Slot Transducer Removal And Installation - Emerson Rosemount 3410 Series Maintenance And Troubleshooting Manual

Gas ultrasonic flow meters
Hide thumbs Also See for Rosemount 3410 Series:
Table of Contents

Advertisement

Maintenance and Troubleshooting Manual
00809-0600-3104
 CAUTION
Escaping gases or fluids hazard
Process Seal Materials Single Seal Certification (T-XX and T-200 Transducers)
Wetted material for T-XX style transducers are 316 stainless steel (SS) or Inconel holders
with Hastelloy-C pins, Stycast 2850 Epoxy, and glass.
Wetted materials for T-200 Style transducers are titanium housing and NBR (Nitrile) or
FKM (Viton) O-ring material.
Only Rosemount specified O-ring replacements shall be used for process seal O-ring
materials for T-200 transducers. No substitutions are allowed to maintain process seal
integrity.
Verify chemical compatibility of material with components of process fluid.
Reference
Failure to select the suitable meter seals may cause escaping gases or liquids, resulting in
injury or equipment damage.
NOTICE
Prior to lifting the meter, refer to the Gas Ultrasonic Flow Meter nameplate or outline
dimensional (general arrangement) drawing for the assembled weight.
3.2 

T-Slot transducer removal and installation

The T-Slot transducer assembly offers improved transducer alignment and superior
acoustic isolation between the transducer and the meter housing.
The net result is improved performance and stability. The assembly is used on Rosemount
3414, 3412 and 3411 meters and is line pressure vented. The gas temperature ranges are
shown in
Table 3-1: Temperature ranges for transducers, mounts, and holders
Transducer type
T-11
T-12
T-21
(1)
T-22
(2)
, T-41
(1)
(1)
T-21 and T-41 transducers use W-01 transformers.
(2)
T-22 transducers use W-01 transformers.
Rosemount Gas Ultrasonic Flow Meters
Parker Seals – Chemical Compatibility Catalog EPS
Table
3-1.
Temperature range
-4 to +212 °F (-20 to +100 °C)
-4 to +212 °F (-20 to +100 °C)
-4 to +212 °F (-20 to +100 °C)
-58 to +212 °F (-50 to +100
°C)
Mount and holder type
Standard mounts/holders,
NBR O-ring Inconel mounts/ 316L Holders,
NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
Standard mounts/Holders,
NBR O-ring Inconel mounts/ 316L Holders,
NBR O-ring Inconel Mounts/Inconel Holders/FKM O-ring
Standard mounts/holders/NBR O-ring
Inconel mounts/316L holders, NBR O-ring
Inconel mounts/Inconel holders/FKM O-ring
Standard mounts/holders/NBR O-ring
Inconel mounts/316L holders, NBR O-ring
Inconel mounts/Inconel holders/FKM O-ring
Meter repairs
January 2024
5350.
39

Advertisement

Table of Contents
loading

This manual is also suitable for:

Rosemount 3414Rosemount 3412Rosemount 3411

Table of Contents